TEC4 is a shielded hot cell with vertical laminar flow over the whole working area. Thanks to its 2 independent workstations it allows two operatores to work simultaneously. TEC4 can be implemented with a motorized lift for the whole hot cell to be moved to the desired height.

Related Prodcuts

µ DDS-A

RSP

Features

Dimensions
External (without motorized lift): 1.930(w) x 930(d) x 2.390(h) mm
External (with motorized lift and related frame): 2.030(w) x 1.000(d) x 2.390(h) + 140(h) mm (lift height range)
Internal: 1.700(w) x 600(d) x 730(h) mm

Shielding

  • Ventilation area 10 mm Pb;
  • Lead glass window, working area, Shielded front doors, Syringe extraction pre-chamber, Waste area 20 mm Pb;
  • Generator pre-chamber, Generator pre-chamber door, Dose calibrator area 40 mm Pb;
  • Additional 10 mm Plexiglas shielding in the working area for Beta emitters

Main features

  • n. 1 GMP Grade A working area - more than 1 square meter
  • n. 2 independent workstations (n. 2 operators)
  • n. 1 air-tight, shielded and vented pre-chamber for TC99 generator housing
  • extractable carousel for up to n.4 TC99 generators, with lifting system for the chosen generator
  • n. 2 air-tight, shielded and vented frontal pre-chambers (double door drawer)
  • in-house dose calibrator shielded area
  • front shielded door with lead glass window
  • shielded housing for waste bin
  • air-tight polycarbonate inner doors with gloves for maintenance and cleaning
  • motorized lift for TEC4 height adjustment (optional)

HMI Panel PC

  • for system control, data saving and display
  • eliminates any other hot cell control device
  • makes the system store, save and trace any event and alarm related to the hot cell work
  • grants multilevel password access
  • makes the system store, save and trace any event, access and alarm related to the hot cell work
  • allows efficient TeleTroubleshooting
  • with smooth design - completely embedded in the external finishing of the cell